Amplitude of the vibration level can be measure with four ways. The
value of amplitude can be measure with peak to peak, zero to peak, RMS and
average. Peak to peak is the distance from the top of the positive peak to the
bottom of the negative peak. This method of measurement is most often used when
referring tp displacement amplitude.
The value of zero to peak is the measurement from to zero line
to the top of the positive peak or the bottom of the negative peak. This type
of the measurement is most often used to describe the vibration level from a
velocity transducer or accelerometer.
The Root Mean Square (RMS) is the true measurement of the power
under the curve. The RMS value is 0.707 x peak only applies to pure sine wave.
The true RMS value can be calculated by the square root of the sum of the squares
of a given number of points under the curve. The formula for true RMS they are:
RMS =
square {(P12 + P22 + P32
+… Pn2)/n}
For measurement the value of the true RMS, the measurement for
signals that it contain pulses must consider the value of the crest factor and
duty cycle. The crest factor (CF) is the ratio of the peak value to the RMS
value with the DC component removed. These formulas can be write:
CF = (P-DC)/RMS
For accurate measurement of pulse, the value of the crest factor
is 7 normally can be used. The duty cycle is the ratio of the pulse width (PW)
to pulse recurrence frequency (PRF). The formula for duty cycle they are:
duty cycle
= PW/PRF
There are also several forms of pseudo are used in some
equipments. This value of the pseudo RMS can be formula with
RMS =
0.707 x peak
RMS =
square (AC2 + DC2 )
Analog meters is the equipments that can be use as to measure average
amplitude. There are various the value of constants are the used to calculate
peak, peak to peak, zero to peak, or RMS. Most measurements that are not true
RMS measurements are either overstated or understated. When describing the
vibration level of a machine, the RMS value should be used if possible. However
some cases require peak to peak measurements, for example when measuring mils
of displacement. Other case require zero to peak displacement measurement such
as high places on the roll.
The average values are
measured by analog meters. The value average is the converted to peak by multiplying
a constant of 1.57. These calculated values are accurate only when measuring
pure sinusoids. The following constants may be helpful. However they apply to
pure sine waves only. The more of the signal deviates from a true sine wave,
the more a error is introduced.
Average
= 0,637 x peak
Average
= 0,90 x RMS
Peak to
peak = 2 x peak
Peak =
1,414 x RMS
Peak =
1,57 x average
RMS =
0,707 x peak
RMS =
1,11 x average
No comments:
Post a Comment